Survival rates in non-small cell lung cancer (NSCLC) are low. Detection of circulating tumor DNA in liquid biopsy (plasma) is increasingly used to identify targeted therapies for clinically actionable mutations, including EGFR mutations in NSCLC. The cobas® EGFR Mutation Test v2 (cobas EGFR test) is FDA-approved for EGFR mutation detection in tissue or liquid biopsy from NSCLC. Standard K2EDTA tubes require plasma separation from blood within 4 to 8 hours; however, Roche Cell-Free DNA (cfDNA) Collection Tubes (Roche cfDNA tube) enable whole blood stability for up to 7 days prior to plasma separation. This analysis assessed performance of Roche cfDNA tubes with the cobas EGFR test for the detection of EGFR mutations in plasma from healthy donors or patients with NSCLC. Overall, test performance was equally robust with either blood collection tube, eg, regarding limit of detection, linearity, and reproducibility, making Roche cfDNA tubes suitable for routine clinical laboratory use in this setting. Importantly, the Roche cfDNA tubes provided more flexibility for specimen handling versus K2EDTA tubes, eg, in terms of tube mixing, plasma separation, and sample stability, and do not require processing of blood within 8 hours thereby increasing the reach of plasma biopsies in NSCLC.
Lung cancer remains a leading cause of cancer-related deaths globally, accounting for 2.2 million new cases and 1.8 million deaths per year [[
Recent advances in the biologic understanding of NSCLC, driven in part by the evolution of advanced diagnostics such as next-generation sequencing (NGS), have shown the importance of the molecular features associated with the disease. Specifically, subsets of NSCLC with specific genetic alterations have been identified, including those that harbor abnormalities in EGFR, ALK, MET, HER2, ROS1, BRAF, RET, NTRK1, PIK3CA, KRAS, and MEK [[
Although molecular analysis of tissue samples is regarded as standard practice in NSCLC, there may be limitations with this approach relating to sampling location, amount of tissue available for sampling, and tumor heterogeneity [[
When collected in standard di-potassium salt of ethylene diamine tetraacetic acid (K2EDTA) blood draw tubes, plasma must be separated from blood within 4 to 8 hours. The Roche Cell-Free DNA (cfDNA) Collection Tube (Roche cfDNA tube [Roche Molecular Systems, Inc., Branchburg, NJ, US) contains an anticoagulant (tri-potassium salt of ethylene diamine tetra acetic acid [K3EDTA]), a cell lysis inhibitor, and a cfDNA stabilizer that allows whole blood to be stable for up to 7 days prior to separation of the plasma. However, it has not been previously reported whether K3EDTA or the cell lysis inhibitor affect the performance of assays designed to detect mutations in ctDNA. The results presented here demonstrate the performance of Roche cfDNA tubes, including a comparison with standard K2EDTA tubes, for the detection of defined EGFR mutations in plasma derived from patients with NSCLC or from healthy donors using the cobas EGFR test.
Between May 2018 through February 2020, plasma samples from healthy donors or patients with stage III or IV NSCLC were derived from whole blood and collected into Roche cfDNA tubes; plasma from patients with NSCLC was co-collected into standard K2EDTA tubes for comparison with Roche cfDNA tubes (8 mL per tube for both tube types). Samples were obtained from commercial vendors, who previously obtained ethical approval for all protocols from their local Institutional Review Board or other appropriate ethics committee. All samples were collected with informed consent. To supplement the samples from patients with EGFR mutation-positive NSCLC, surrogate samples were prepared using whole blood from healthy EGFR mutation-negative donors spiked with sheared cell-line DNA to approximately 1.5× the limit of detection (LoD) of the corresponding target sequence detected and reported by the cobas EGFR test. The predominant EGFR mutation for each group (eg, exon 19 deletion [Ex19Del], L858R, G719X, and exon 20 insertion [Ex20Ins]) reported by the cobas EGFR test was used for each surrogate sample (Table 1). Cell-line DNA was procured from a commercial vendor (Horizon Discovery, Cambridge, UK), sheared, and characterized, eg, by calibration of ctDNA with a known quantitation of target mutation sequences, prior to use. For the clinical LoD confirmation analysis, the concentration of EGFR mutations present in NSCLC samples was similarly estimated to allow formulation of the panel at approximately 1× LoD. The mutation status of the samples from patients with NSCLC was confirmed by a validated in-house NGS test using the Illumina MiSeq System (Illumina, Inc., San Diego, CA, US).
Graph
Table 1 The predominant EGFR mutation for each group reported by the cobas EGFR test.
Exon COSMIC ID Exon 18 G719X 2156G>A 6252 Exon 19 Ex19Del ex19del 2235_2249del15 6223 Exon 20 S768I 2303G>T 6241 T790M 2369C>T 6240 Ex20Ins 2307_2308insGCCAGCGTG 12376 Exon 21 L858R 2573T>G 6224 L861Q 2582T>A 6213
1
2 COSMIC ID, Catalogue Of Somatic Mutations In Cancer identification; Ex19Del, exon 19 deletion; Ex20Ins, exon 20 insertion.
The cobas EGFR test is a real-time polymerase chain reaction (PCR) assay consisting of two major processes: (
Plasma from patients with NSCLC and healthy donors was tested using a validated, targeted DNA amplicon sequencing NGS technology to confirm detected EGFR mutations. For each NSCLC and healthy donor sample collected in K2EDTA tubes, PCR-based DNA input was used to amplify regions in each exon, followed by dual-indexing PCR. The sample libraries were pooled and sequenced along with positive controls, using the 2× 150 base pair paired-end sequencing protocol on the Illumina MiSeq System.
Samples for an interference study utilized plasma from whole blood from healthy donors collected with one lot of Roche cfDNA tubes. Sheared EGFR mutation-positive cell-line DNA was diluted into four sets of the collected plasma at 2× LoD, along with one set containing no EGFR mutation-positive cell-line DNA. The plasma was then spiked with potential interfering substances and tested along with control conditions (no interfering substances) and assessed with the cobas EGFR test. The following sample types were evaluated: control (no potentially interfering substance [with and without diluents used for substances, eg, normal saline or H
The test concentrations for hemoglobin, triglycerides, albumin, and bilirubin were from the Clinical and Laboratory Standards Institute (CLSI) EP07-A2 guideline [[
Roche cfDNA tube performance was evaluated against predefined acceptance criteria. LoD verification was defined as the proportion of correct results as a percentage of the total number of valid results. For clinical LoD verification and the reproducibility study, data were summarized by the percentage agreement by panel member with the associated 95% exact confidence intervals. Linearity was evaluated according to CLSI document EP06-A [[
In the method comparison study, the positive, negative, and overall percentage agreement (OPA) between the EGFR mutation results from plasma collected in Roche cfDNA tubes and plasma collected in K2EDTA tubes were calculated, with 95% confidence intervals also reported. The specimen stability study compared the results for each sample at each time point with the corresponding Time 0 result. The last time point for a given storage condition at and before which the results agreed was determined to be the stability for the storage condition. The mixing study compared the observed results for each test condition with the expected results. If the results agreed, the condition passed. The centrifugation study compared the observed results for each test condition with the expected results. If the results agreed, the condition passed. The surrogate sample study used a hit rate comparison to a reference panel using P-values from the Fisher's exact test for levels that had less than a 100% hit rate. The interfering substances study compared the observed results in the presence and absence of potentially interfering substances or material with the expected results. If the results agreed, the potentially interfering substance or material was determined to not interfere with the cobas EGFR test.
A comparison was made using EGFR mutation-positive samples (Ex19Del, L858R, and T790M) diluted from approximately 2× LoD to 0.03× LoD, along with an EGFR mutation-negative control, and four panels: 1) sheared cell-line DNA diluted into healthy donor plasma; 2) sheared cell-line DNA diluted into EGFR mutation-negative NSCLC plasma; 3) sheared cell-line DNA diluted into healthy donor whole blood prior to plasma separation; and 4) EGFR mutation-positive plasma diluted in EGFR mutation-negative NSCLC patient plasma samples (reference panel). All mutations were confirmed using sequencing.
Twenty replicates from each panel were tested across two lots of the cobas EGFR test reagents. The hit rates for each panel level and mutation were compared with the reference panel (S1–S3 Tables); all levels had a 100% hit rate or a P value ≥0.05, except for two panel levels for Ex19Del (approximately 0.03× LoD and 0.13× LoD in Panels 2 and 3, respectively). Based on this analysis, the results indicate that the surrogate samples perform similarly to the EGFR mutation-positive NSCLC plasma reference (Condition 4).
This study evaluated the minimum number of inversions needed to mix whole blood with anticoagulant after collection in the Roche cfDNA tube, and aimed to determine whether increased exposure to the Roche cfDNA tube rubber stopper through multiple inversions would impact the performance of the cobas EGFR test.
One lot of Roche cfDNA tubes was utilized to collect whole blood from healthy donors. After collection, each tube was gently inverted a specified number of times by hand. Testing was performed using surrogate samples of sheared EGFR mutation-positive cell-line DNA diluted into the whole blood at 2× LoD and mixed according to the specified number of inversions in S4 Table. Cell-line DNA was not added to the EGFR mutation-negative samples.
The performance of the cobas EGFR test was not affected by any of the conditions. A minimum of four total inversions was needed to mix the blood collected in the Roche cfDNA tube, as fewer than four inversions resulted in increased hemolysis (S1 Fig). There was also no impact to the cobas EGFR test results for G719X, Ex19Del, S768I, T790M, Ex20Ins, L858R, and L861Q mutations after increased exposure (30 inversions) to the tube rubber stopper.
A guard band study was performed to determine whether deviations from the recommended centrifugation operating range of ≤1600 x g for 10 to 15 minutes influenced the performance of the cobas EGFR test.
Samples for this study utilized whole blood from healthy donors collected with one lot of Roche cfDNA tubes. Sheared EGFR mutation-positive cell-line DNA was diluted into four sets of the collected whole blood at 2× LoD, along with one set containing no EGFR mutation-positive cell-line DNA. The collected whole blood was then centrifuged at various times and speeds (S5 Table), and the resulting plasma assessed with the cobas EGFR test. All results gave the correct mutation call for all tested samples (Ex19Del, Ex20Ins, S768I, G719A, L861Q, T790M, L858R, and wildtype) for each centrifugation condition. The tested centrifugation conditions had no impact on the performance of the cobas EGFR test.
The LoD, previously established with K2EDTA plasma, for the cobas EGFR test was verified using plasma from Roche cfDNA tubes for exon 18 G719X mutations, Ex19Del mutations, exon 20 S768I, T790M, Ex20Ins mutations, and exon 21 L858R and L861Q mutations. Twenty replicates were tested per mutation using surrogate samples and one batch of cobas EGFR test reagents. The LoD for EGFR mutation detection in Roche cfDNA tubes' plasma was verified as ≤100 cp/mL, as shown in Table 2.
Graph
Table 2 Verification of LoD for EGFR mutation detection.
Test level (cp/mL) Valid results ( Correct results ( Hit rate (%) 95% CI Ex19Del 75 20 20 100 83.2, 100 S768I 25 20 20 100 83.2, 100 L858R 100 20 20 100 83.2, 100 T790M 100 20 20 100 83.2, 100 L861Q 30 20 20 100 83.2, 100 G719X 100 20 20 100 83.2, 100 Ex20Ins 25 20 20 100 83.2, 100
3 CI, confidence interval; cp, copies; Ex19Del, exon 19 deletion; Ex20Ins, exon 20 insertion; LoD, limit of detection.
This analysis aimed to verify the sensitivity of the cobas EGFR test using samples from patients with NSCLC, collected in Roche cfDNA tubes. Two panels of EGFR mutation-positive NSCLC specimens for Ex19Del, L858R, T790M, and S768I were diluted to approximately the LoD for each mutation group in pooled NSCLC EGFR mutation-negative plasma collected in two Roche cfDNA tube lots. Three unique Ex19Del mutations were evaluated, the two most predominant plus one additional mutation. EGFR mutation-negative NSCLC samples were tested in plasma from both lots of Roche cfDNA tubes. All mutations were confirmed by NGS. Panels were tested on 2 non-consecutive days by two operators at each of three testing sites using one lot of cobas EGFR test reagents. Table 3 shows the mutations tested, the target concentrations, and the OPA for each mutation; all mutations had an OPA of ≥95.8%.
Graph
Table 3 EGFR mutations tested, target concentrations, and OPA for each EGFR mutation.
Concentration (cp/mL) Valid tests ( Agreement ( OPA, % (95% CI) Wildtype – 48 48 100 (92.6, 100) Ex19del 2235_2249del15 75 48 48 100 (92.6, 100) Ex19del 2236_2250del15 75 48 48 100 (92.6, 100) Ex19del 2236_2253dup TTCCCGTCGCTATCAAGG 75 48 48 100 (92.6, 100) L858R 100 48 47 97.9 (88.9, 99.9) T790M 100 48 47 97.9 (88.9, 99.9) S768I 25 48 46 95.8 (85.7, 99.5)
4 CI, confidence interval; cp, copies; Ex19Del, exon 19 deletion; OPA, overall percentage agreement.
Linearity of the cobas EGFR test, established in K2EDTA plasma, was verified using plasma from Roche cfDNA tubes. Surrogate samples were used to test each mutation group using two replicates at 1.0 × 10
All mutations met the requirements, and the linearity for the ranges shown in S6 Table was verified. The linear fit and the best second- or third-order polynomial fit are shown in Fig 1, and the differences between the regression analyses are shown in S7–S13 Tables.
Graph: Linear fit and the best second- or third-order polynomial fit for the surrogate samples for EGFR mutations: A: Ex19Del; B: S768I; C: L858R; D: T790M; E: L861Q; F: G719X; and G: Ex20Ins. Ex19Del, exon 19 deletion; Ex20Ins, exon 20 insertion; SQI, Semi-Quantitative Index.
The diagnostic accuracy of the cobas EGFR test using plasma from Roche cfDNA tubes was assessed by comparing results of EGFR mutation-positive NSCLC samples collected in Roche cfDNA tubes with samples collected in K2EDTA tubes simultaneously from the same donor.
A total of 17 patients with EGFR mutation-positive NSCLC and 34 patients with NSCLC who were mutation-negative for EGFR in plasma were included in this study. Only EGFR-positive mutations with at least one mutation that met a minimum requirement (S14 Table) to be consistently detectable at approximately 1× LoD were used in the analysis.
A total of 71 samples (51 from patients with NSCLC [mutation-positive NSCLC: 17 patients; EGFR mutation-negative in plasma: 34 patients] and 20 surrogate samples) were tested in plasma from both the Roche cfDNA and K2EDTA tubes. There was 100% agreement for all samples, as shown in Tables 4 and 5. Some samples contained more than one EGFR mutation, resulting in more than 71 total results. The agreement for each mutation is shown in Table 6.
Graph
Table 4 Agreement between the Roche cfDNA tubes and K2EDTA tubes.
K2EDTA tubes MD NMD Total Roche cfDNA tubes MD 37 0 37 NMD 0 34 34 Total 37 34 71
5 cfDNA, cell-free DNA; MD, mutation detected; NMD, no mutation detected.
Graph
Table 5 Overall agreement between all samples.
Agreement Valid results ( Correct results ( Hit rate (%) 95% CI Positive 37 37 100 90.5, 100 Negative 34 34 100 89.7, 100 Overall 71 71 100 94.9, 100
6 CI, confidence interval.
Graph
Table 6 Agreement between K2EDTA tubes and the Roche cfDNA tubes for each EGFR mutation.
K2EDTA tubes Ex19Del S768I L858R T790M L861Q G719X Ex20Ins Wildtype Total Roche cfDNA tubes Ex19Del 15 – – – – – – – 15 S768I – 2 – – – – – – 2 L858R – – 16 – – – – – 16 T790M – – – 16 – – – – 16 L861Q – – – – 2 – – – 2 G719X – – – – – 2 – – 2 Ex20Ins – – – – – – 2 – 2 Wildtype – – – – – – – 34 34 Total 15 2 16 16 2 2 2 34 89
- 7 Dashed lines indicate that samples were not positive for the mutation detected.
- 8 cfDNA, cell-free DNA; Ex19Del, exon 19 deletion; Ex20Ins, exon 20 insertion.
Reproducibility of the Roche cfDNA tubes with the cobas EGFR test was evaluated with three lots of tubes, three testing sites, two operators per site, and on 3 non-consecutive days. One lot of the cobas EGFR test reagents was used. Three 9-member panels were made using surrogate samples of sheared EGFR mutation-positive cell-line DNA diluted into healthy donor plasma collected in three unique Roche cfDNA tube lots. Each mutation group was diluted to approximately 100 cp/mL and 300 cp/mL. Samples included mutation-negative, an exon 18 mutation (G719A), an Ex19Del, three exon 20 mutations (S768I, T790M, and Ex20Ins), and two exon 21 mutations (L858R and L861Q), some of which were combined to make individual panel members. Each panel was confirmed by sequencing.
The OPA for each mutation member is shown in Table 7. All panel members had a percentage agreement of ≥98.6%. The overall coefficient of variation for the cycle threshold from valid mutation members ranged from 4.4% to 13.7% across all mutation members (S15 Table). Within each component, coefficient of variation ranged from 0.0% to 10.4% across all mutation members.
Graph
Table 7 OPA for each mutation member.
Panel member–level Valid tests ( Agreement ( OPA, % (95% CI) Wildtype–N/A 72 71 98.6 (92.5, 100) G719A – 100 cp/mL 72 71 98.6 (92.5, 100) Ex19Del– 100 cp/mL 72 72 100 (95.0, 100) Ex20Ins– 100 cp/mL 72 72 100 (95.0, 100) S768I – 100 cp/mL 72 72 100 (95.0, 100) T790M – 100 cp/mL 72 72 100 (95.0, 100) L858R – 100 cp/mL 72 72 100 (95.0, 100) L861Q – 100 cp/mL 72 72 100 (95.0, 100) G719A – 300 cp/mL 72 72 100 (95.0, 100) Ex19Del– 300 cp/mL 71 71 100 (94.9, 100) Ex20Ins– 300 cp/mL 71 71 100 (94.9, 100) S768I – 300 cp/mL 72 72 100 (95.0, 100) T790M – 300 cp/mL 71 71 100 (94.9, 100) L858R – 300 cp/mL 72 72 100 (95.0, 100) L861Q – 300 cp/mL 71 71 100 (94.9, 100)
9 CI, confidence interval; cp, copies; Ex19Del, exon 19 deletion; Ex20Ins, exon 20 insertion; N/A, not applicable; OPA, overall percentage agreement.
An interference study was performed to determine whether high levels of hemoglobin, triglycerides, bilirubin, or albumin, or the preservative solution in plasma from the Roche cfDNA tubes, would interfere or influence the performance of the cobas EGFR test.
All results gave the correct mutation call for all tested samples (Ex19Del, Ex20Ins, S768I, G719A, L861Q, T790M, L858R, and wildtype) for each condition. The tested potential interfering substances had no impact on the performance of the cobas EGFR test.
The stability and storage conditions of whole blood and plasma obtained from patients with NSCLC from the Roche cfDNA tubes were evaluated with the cobas EGFR test. Stability of NSCLC specimens was evaluated using five conditions, which included at least one wildtype and five EGFR mutation-positive samples. For each condition, one replicate per patient was tested at Time 0 and each time point: time (T0) (no storage of whole blood); Condition 1 (storage of whole blood for 25 hours at 32°C followed by storage at 27°C for 4 days [T1], 7 days [T2], and 8 days [T3]); Condition 2 (storage of plasma at 32°C for 5 hours [T4] and 25 hours [T5]); Condition 3 (storage of plasma at 2°C to 8°C for 4 days [T6] and 8 days [T7]; Condition 4 (storage of plasma at –25°C to –15°C for 31 days [T8] and 13 months [T9] with 3× freeze/thaw); and Condition 5 (storage of plasma at ≤–70°C for 31 days [T10] and 13 months [T11] with 3× freeze/thaw.
Results from the study were compared with T0 for each condition (Conditions 1 to 5; Table 8). Initially, Condition 1 was tested in comparison with T0. Once the whole blood storage condition was determined (T3), the subsequent testing of Conditions 2, 3, 4, and 5 followed Condition 1 T3 to separate the plasma prior to evaluating plasma storage. As a consequence, the NSCLC specimens used for each condition varied; however, the specimens tested for a given condition and its corresponding T0 were the same. EGFR wildtype and mutation-positive results from each patient at each time point had the same results from the corresponding T0 EGFR mutation-negative and mutation-positive samples. Results from the study further showed that whole blood collected in Roche cfDNA tubes can be stored for up to 25 hours at 32°C followed by up to 8 days at 27°C (Condition 1, T3), and following whole blood storage indicated above, plasma may be stored for up to: 1) 25 hours at 32°C (Condition 2, T5); 2) 8 days at 2°C to 8°C (Condition 3, T7); 3) 13 months at –25°C to –15°C (Condition 4, T9) with 3× freeze/thaw; and 4) 13 months at ≤–70°C (Condition 5, T11) with 3× freeze/thaw.
Graph
Table 8 Summary of specimen handling stability results.
Condition Time point Temperature Time Pass/Fail C1 (whole blood) T1 32°C + 27°C 25 hours + 4 days Pass T2 32°C + 27°C 25 hours + 7 days Pass T3 32°C + 27°C 25 hours + 8 days Pass C2 T4 32°C 5 hours Pass T5 32°C 25 hours Pass C3 T6 2°C to 8°C 4 days Pass T7 2°C to 8°C 8 days Pass C4 T8 –25°C to –15°C 31 days Pass T9 –25°C to –15°C 13 months Pass C5 T10 ≤–70°C 31 days Pass T11 ≤–70°C 13 months Pass
10
The detection of ctDNA in liquid biopsy (plasma), has become increasingly used to detect clinically actionable mutations in patients with cancer. However, DNA has a short half-life in plasma and can be affected by the lysis of white blood cells present in blood, which increases the background DNA and thus dilutes the ctDNA present. Plasma must be separated from whole blood within 4 to 8 hours to prevent these challenges when blood is collected in standard K2EDTA blood collection tubes. In comparison, the Roche cfDNA tubes contain an anticoagulant (K3EDTA), as well as an inhibitor of cell lysis, which reduces the urgency for plasma separation, allowing laboratories greater flexibility in transportation and time to processing of blood for ctDNA detection. Using the Roche cfDNA tubes, blood specimens are stable for 7 days when stored or shipped between 15°C and 25°C, with transient excursions of up to 16 hours to 15°C to 30°C [[
Using the cobas EGFR test, which is FDA-approved for detection of EGFR mutations in plasma collected in K2EDTA tubes from patients with NSCLC [[
This analysis also found that the performance of the cobas EGFR test was equally robust with either type of blood collection tube, eg, with regards to LoD, linearity, and reproducibility [[
The cobas EGFR test was previously shown to have robust accuracy and reproducibility in both preclinical and clinical studies using plasma collected in standard EDTA anticoagulation blood draw tubes. This study shows that the cobas EGFR test is similarly robust when using plasma obtained in Roche cfDNA tubes for up to 7 days, indicating that cfDNA remains detectable within this timeframe, thus making them suitable for routine clinical laboratory use in this setting. Furthermore, due to the preservative solution, the Roche cfDNA tubes enable draws that are not dependent upon processing within 8 hours, therefore increasing the reach of plasma biopsies for patients with NSCLC.
S1 Fig
Representative plasma image for mixing by inversion.
S1 Table
Comparison between surrogate samples and reference panel for Ex19Del (n = 20).
S2 Table
Comparison between surrogate samples and reference panel for L858R (n = 20).
S3 Table
Comparison between surrogate samples and reference panel for T790M (n = 20).
S4 Table
Inversions for mixing.
S5 Table
Centrifugation conditions.
Samples used in the study included Ex19Del, EX20Ins, S768I, G719A, L861Q, T790M, L858R, and wildtype.
S6 Table
Concentration range for each EGFR mutation group.
cp, copies; Ex19Del, exon 19 deletion; Ex20Ins, exon 20 insertion.(DOCX)
S7 Table
Predicted SQI from regression analysis for Ex19Del.
cp, copies; Ex19Del, exon 19 deletion; SD, standard deviation; SQI, Semi-Quantitative Index.(DOCX)
S8 Table
Predicted SQI from regression analysis for S768I.
cp, copies; SD, standard deviation; SQI, Semi-Quantitative Index.(DOCX)
S9 Table
Predicted SQI from regression analysis for L858R.
S10 Table
Predicted SQI from regression analysis for T790M.
cp, copies; SD, standard deviation; SQI, Semi-Quantitative Index.(DOCX)
S11 Table
Predicted SQI from regression analysis for L861Q.
S12 Table
Predicted SQI from regression analysis for G719X.
cp, copies; N/A, not applicable; SD, standard deviation; SQI, Semi-Quantitative Index.(DOCX)
S13 Table
Predicted SQI from regression analysis for Ex20Ins.
S14 Table
Minimum SQI required for EGFR mutations.
S15 Table
Overall CV for the cycle threshold from valid mutation members.
cp, copies; CV, coefficient of variation; Ex19Del, exon 19 deletion; Ex20Ins, exon 20 insertion.(DOCX)
We would like to thank Birgit Mortensen (Aarhus University Hospital, Aarhus, Denmark) for her skillful technical assistance. Research support for third-party writing assistance for this manuscript, furnished by Stephen Salem, BSc, of Health Interactions. Additional editorial support was provided by Estelle Challinor of inScience Communications (Manchester, UK). COBAS is a trademark of Roche. All other product names and trademarks are the property of their respective owners.