Development of the Characterization Methods Without Electrothermal Feedback for TES Bolometers for CMB Measurements

Nishinomiya, Yume ; Kusaka, Akito ; et al.
In: Journal of Low Temperature Physics, Jg. 209 (2022-09-19), S. 1079-1087
Online unknown

Development of the Characterization Methods Without Electrothermal Feedback for TES Bolometers for CMB Measurements 

Superconducting Transition-Edge Sensor (TES) bolometers are used for cosmic microwave background (CMB) observations. We used a testbed to evaluate the thermal performance of TES bolometers in regard to the saturation power P sat and intrinsic thermal time constant τ 0 . We developed an evaluation method that is complementary to methods with electrothermal feedback. In our method, the antenna termination resistor of the bolometer is directly biased with DC or AC electric power to simulate optical power, and the TES is biased with small power, which allows P sat and τ 0 to be determined without contribution from the negative electrothermal feedback. We describe the method and results of the measurement using it. We evaluated P sat of five samples by applying DC power and confirmed the overall trend between P sat and the inverse leg length. We evaluated τ 0 of the samples by applying DC plus AC power, and the measured value was reasonable in consideration of the expected values of other TES parameters. This evaluation method enables us to verify whether a TES has been fabricated with the designed values and to provide feedback for fabrication for future CMB observations.

Keywords: Bolometer; Cosmic microwave background; Electrothermal feedback; Intrinsic time constant; Saturation power; Transition-edge sensor

Copyright comment Springer Nature or its licensor holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.

Introduction

A Transition-Edge Sensor (TES) [[1]] has a narrow temperature region between normal and superconducting states and is used as a thermometer in a bolometer, which can measure the power of incoming photons. TES bolometers have been adopted in cosmic microwave background (CMB) observations due to their supreme sensitivity [[2]–[5]].

In this paper, we focus on evaluation of two relevant TES parameters: the saturation power Psat and intrinsic thermal time constant τ0 . Psat is the amount of electrical power required to drive the TES normal in the absence of absorbed optical power. Its desired value is informed by the optical power from the sky absorbed by the TES bolometers [[6]]. Psat is related to TES thermal carrier noise due to fluctuations in heat flow between TES and the heat bath as NEPgPsat , where NEPg is the noise equivalent power of the thermal carrier noise [[7]]. τ0 is defined as C/G, which is the ratio of the heat capacity of the bolometer island C and the thermal conductivity G between the TES and the heat bath. The response speed of TES is faster than τ0 due to negative electrothermal feedback. The time constant with the feedback can be defined as τeff=τ0/(L+1) , where L is the loop gain related to the TES bias power. Since τeff acts as a low-pass filter for the TES current, the requirement of τeff for CMB experiments is set based on the sampling rate and signal modulation frequency [[8]] in addition to stability.

We developed an evaluation method for Psat and τ0 without considering electrothermal feedback. Our evaluation method enables us to provide feedback to the fabrication. These parameters are tuned in the fabrication process of the TES to satisfy the observation requirements [[9]]. In the fabrication, Psat is tuned by varying the length of the bolometer legs because it is proportional to G, while τ0 is tuned by varying C. Our method is complementary to I-V measurement with the feedback [[8], [10]], in which the TES current is measured while sweeping the bias voltage. We can measure τ0 directly with our method, in contrast to the I-V method where τ0 is estimated from τeff as a function of the loop gain L [[8]]. Here, we describe the methods and results of the measurement using them.

Saturation Power Measurement

Setup

We measured TES test samples for the development of a next-generation ground-based CMB experiment, provided by UC Berkeley. The samples have an antenna termination resistor coupled with a TES thermally as shown in Fig. 1(left). In actual operation, optical signals from the antenna travel on microstrip transmission lines, and the termination resistor converts the optical signals into thermal signals. Here, instead we applied DC electric power to the termination resistor to simulate an optical signal.

Figure 1(right) shows the circuit of the measurement setup. The bolometer sample was installed on the mixing chamber stage of a dilution refrigerator.[1] The stage temperature was controlled by a heater on the stage under closed-loop PID control with an accuracy of around 0.1 mK. Electric power to the termination resistor was provided by a room-temperature DC power supply. The TES was biased with a small AC current with frequency of 9.8 Hz with an AC resistance bridge.[2] We measured the TES resistance by recording the TES voltage. The bias power ( 0.01 pW) was much smaller than Psat ( 1-10 pW), so Psat could be determined without electrothermal feedback.

Graph: Fig. 1 Left: the photograph of the measured bolometer sample. The antenna termination resistor is thermally coupled with the TES. Right: the circuit of the measurement setup. TES was biased with small AC power. An oscilloscope was used only in the time constant measurement (Sect. 3). A DC power supply (Sect. 2) or a signal generator (Sect. 3) provided electric power to the termination resistor (Color figure online)

Evaluation Method and Results

We defined Psat as equivalent to Pbath(Tc) , which is the power flow to the heat bath when the TES island temperature is at critical temperature Tc . This can be expressed using the bath temperature Tb , index n, and G [[1]]:

  • Psat=Pbath(Tc)=K(Tcn-Tbn)=GnTcn-1(Tcn-Tbn)
  • Graph

    where the thermal conductivity at Tc is well defined using constant K as G=dPbath(T)/dT|Tc=nKTcn-1 . At steady state, Pbath=Popt+Pbias , where Popt is the optical power TES receives, and Pbias is the TES bias power. In our measurements, Popt was replaced with the DC electric power dissipated in the resistor, Pres ( 1-10 pW), and Pbias ( 0.01 pW) could be neglected. Therefore, we assumed that Pbath=Pres .

    First, we measured the TES resistance R while sweeping Tb without applying Pres and then determined Tc from the R- Tb curve.[3] The small bias power in our method enables the measurement of Tc at zero current; increasing the current may shift and widen the transition [[11]]. Next, we applied DC power to the termination resistor and repeated the R- Tb curve measurement. Figure 2(left) shows the result of the R- Tb curves of one bolometer for various Pres . To consider systematic errors, the error bars were re-scaled so that the reduced chi square reached 1. We found that the transition occurred at lower Tb when higher Pres was applied. From the R- Tb curves, we determined Tb,trans , which is Tb in transition with a given Pres . Figure 2(right) shows the relation between Pres and Tb,trans of five TES samples with different leg length. The data are fitted with Eq. (1) to estimate G and n. Then we determined Psat at the nominal operating temperature of 100 mK.

    Graph: Fig. 2 Left: the measured resistance of one sample (100 μm leg length) as a function of bath temperature for various Pres. The error bars were calculated by the standard deviation from multiple measurements and the systematics. Right: the relation between Pres and Tb,trans of five samples with different leg length ℓ. The lines show fitting results with Eq. (1). Psat at 100 mK was estimated from the fitting lines as the red vertical line shows (Color figure online)

    Graph: Fig. 3 The measured Psat as a function of the inverse leg length. The dashed line is a linear fit (Color figure online)

    The results of G, n, and Psat are shown in Table 1. All estimated n values of the five samples were around 3, which is reasonable because n is the thermal conductance index and was expected to be between 2 and 4 due to the material's properties [[7]]. Psat was expected to be inversely proportional to the leg length [[12]]. As shown in Fig. 3, we confirmed the overall trend between Psat and the inverse leg length, although the linear fit result was influenced by possible systematic effects, such as the extra systematic errors in Psat fitting and slight geometrical variation of the five samples.

    Table 1 The estimated parameters of five TES samples with different leg lengths

    ID

    [μm]

    Tc [mK]

    G [pW/K]

    n

    Psat(100 mK) [pW]

    1

    500

    187.7 ± 0.2

    53 ± 4

    2.9 ± 0.4

    2.86 ± 0.06

    2

    200

    186.6 ± 0.2

    103 ± 1

    2.93 ± 0.03

    5.51 ± 0.04

    3

    130

    185.7 ± 0.2

    129 ± 4

    2.9 ± 0.2

    6.9 ± 0.1

    4

    100

    192.1 ± 0.1

    172 ± 1

    3.03 ± 0.03

    9.38 ± 0.03

    5

    50

    190.0 ± 0.3

    248 ± 1

    3.20 ± 0.03

    12.86 ± 0.03

    The error of Tc comes from fitting of R- Tb curve. The errors of G and n come from fitting of Pres - Tb,trans curve. Psat at 100 mK was determined using G and n

    Time Constant Measurement

    Setup

    To measure the intrinsic time constant τ0 , we applied AC electric power to the termination resistor, and measured the phase shift of the TES output voltage that was modulated by the electric power. The measurement setup was almost the same as that of Psat measurement as described in Sect. 2, except for two differences. Instead of a DC power supply, we used a signal generator to provide differential signals consisting of DC power plus a small amount of sinuous power. The AC resistance bridge can output a raw analog signal of TES voltage as well as a digital signal, so we measured the analog output signal using an oscilloscope.

    Evaluation Method and Results

    The TES thermal differential equation is shown below [[1]]:

    2 CdTdt=-Pbath+Pbias+Popt

    Graph

    where T is the TES temperature. We considered a case where the TES is in transition with a small signal limit and Pbath=Pbath0+GΔT , where Pbath0 is Pbath at steady-state. In our measurement Pbias ( 0.1 pW) is negligible. Popt is replaced with a DC offset plus small sinuous electric power with fres applied to the termination resistor, (Popt=) P0+ΔP(t)=P0+Prescos(2πfrest) . We can rewrite Eq. (2) and solve the differential equation for ΔT(t) :

    3 ΔT(t)=PresG11+(2πfresτ0)2cos2πfrest+ϕ(fres)

    Graph

    where

    4 ϕ(fres)=-arctan(2πfresτ0)

    Graph

    is the phase shift due to the electric power and τ0=C/G . We can write the TES resistance R(t) as Rbias(1+αΔT/Tc) , neglecting the TES current term. When we bias TES with the AC current I(t)=Ibiascos(2πfTESt) , the TES output voltage V=R(t)I(t) is:

    5 V=RbiasIbiascos(2πfTESt)+Acos(2π(fres+fTES)t+ϕ)+cos(2π(fres-fTES)t+ϕ)

    Graph

    where A=αPres/(2GTc1+(2πfresτ0)2) . We measured ϕ from the TES output voltage (Eq. (5)) and estimated τ0 from Eq. (4).

    We set P0= 6.3 pW, which almost corresponds to the measured Psat , and Pres= 0.15 pW, which is 2% of Psat . We swept fres from 10 to 50 Hz for 4 hours, over which the bath temperature of 100 mK and the TES resistance on transition were kept stable. We measured the TES voltage with an oscilloscope over 64 seconds for each fres with a sampling rate of 3 kHz. Figure 4 shows an example of the time-ordered output signals with fres of 16 Hz and fTES of 9.8 Hz.

    Graph: Fig. 4 The example of the time-ordered signals (fres=16 Hz, fTES=9.8 Hz). The blue line is the TES output signal amplified by 106, orange is the reference square signal of Pbias at fTES, and green is the electric power including the DC offset and small sinuous wave at fres (Color figure online)

    Graph: Fig. 5 The amplitude of the Fourier transform of TES output signal (fres=16 Hz, fTES=9.8 Hz). The square and triangle mark correspond to the components of frequency (fres±fTES) (Color figure online)

    Graph: Fig. 6 Left: the phase of fres±fTES components measured at various Pbias (0.01 or 0.1 pW) and fTES (9.8 or 18.2 Hz). Note that the x-axis is the actual frequency, not fres. The dashed line is the phase of the measurement setup. Right: the phase after subtraction of the phase of the measurement setup as a function of fres. The solid line is a fit result (Color figure online)

    The analysis procedure consisted of four steps: (I) A template signal was made by mixing the resistor signal (Fig. 4green) with the TES input square signal, which had an amplitude of 5V TTL (Fig. 4orange) in the time domain. The TES input square signal was provided by the AC resistance bridge and synchronized to fTES , so the template signal had two components proportional to cos(2π(fres±fTES)t) . (II) To extract the information of the modulated waves from the TES output signal, cos(2π(fres±fTES)t) , the Fourier transform of the TES output signal (Fig. 5) was cross-correlated with the Fourier transform of the template from step (I). (III) The phase of the modulated waves was computed, and (IV) the phase delay due to the measurement setup, including the preamplifier and wirings, was subtracted as shown in Fig. 6.

    The phases are consistent after subtraction of that of the measurement setup in all measurement conditions (Fig. 6right), even though we measured the phase shift with different fres and TES bias power. The data are fitted with Eq. (4) plus a DC offset ϕ0 : ϕ=-arctan(2πfresτ0)+ϕ0 . τ0 of the TES sample was estimated as 3.0 ± 0.1 msec, while ϕ0 was estimated as -4.6 ± 0.8 deg. The systematic error of τ0 was estimated to be 0.6 msec, which is the difference of the fits between free ϕ0 and fixed ϕ0=0 . The measured value of τ0 is consistent with the expected value within a factor of 2, considering the estimated value of C (0.28 pJ/K) from the specific heat of the normal metal on the bolometer island [[14]] and the measured value of G (129 pW/K) with the method described in Sect. 2.

    Conclusion

    We demonstrated evaluation methods for TES parameters by applying DC or AC electric power to simulate optical power. These methods enabled us to evaluate Psat and τ0 without electrothermal feedback because only small TES bias power is applied. These methods are necessary to characterize TES samples and to provide feedback for the fabrication and development of TES bolometers for future CMB experiments [[15]].

    Acknowledgements

    This work was supported by JSPS KAKENHI Grant Number JP18H05539, JP19H04608, JP19K14736, JP19K21873. Y.N. and T.T. acknowledge the support by FoPM, WINGS Program, the University of Tokyo.

    Publisher's Note

    Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

    References 1 K. Irwin, G. Hilton, Top. Appl. Phys. 99, 63–150 (2005) 2 Adachi S. Astrophys. J. 2020; 897; 1: 55. 10.3847/1538-4357/ab8f24 3 Choi SK. J. Cosmol. Astropart. Phys. 2020; 2020; 12: 045. 10.1088/1475-7516/2020/12/045 4 Sayre J. Phys. Rev. D. 2020; 101; 12. 10.1103/PhysRevD.101.122003 5 Ade P. Phys. Rev. Lett. 2021; 127; 15. 10.1103/PhysRevLett.127.151301 6 C.A. Hill, et al, in Millimeter, Submillimeter, and Far-Infrared Detectors and Instrumentation for Astronomy IX, vol. 10708, pp. 698–718 (2018) 7 Mather JC. Appl. Opt. 1982; 21; 6: 1125. 10.1364/AO.21.001125 8 N.F. Cothard, et al, in Millimeter, Submillimeter, and Far-Infrared Detectors and Instrumentation for Astronomy X, vol. 11453, p. 1145325 (2020) 9 Westbrook B. J. Low Temp. Phys. 2018; 193; 5: 758. 10.1007/s10909-018-2059-0 J.R. Stevens, et al, J. Low Temp. Phys, pp. 1–9 (2020) Bennett DA. J. Low Temp. Phys. 2012; 167: 102. 10.1007/s10909-011-0431-4 A. Suzuki, Multichroic bolometric detector architecture for cosmic microwave background polarimetry experiments. Ph.D. thesis, UC Berkeley (2013) Koopman BJ. J. Low Temp. Phys. 2018; 193; 5: 1103. 10.1007/s10909-018-1957-5 Rayne JA. Phys. Rev. 1957; 107; 3: 669. 10.1103/PhysRev.107.669 K.N. Abazajian, et al, arXiv preprint arXiv:1610.02743 (2016) Footnotes Oxford Io, https://nanoscience.oxinst.com/ LakeShore372, https://www.lakeshore.com/ In I-V measurement, Tc is estimated from a fit of the Psat - Tbath curves [[10]].

    By Yume Nishinomiya; Akito Kusaka; Kenji Kiuchi; Tomoki Terasaki; Johannes Hubmayr; Adrian Lee; Heather McCarrick; Aritoki Suzuki and Benjamin Westbrook

    Reported by Author; Author; Author; Author; Author; Author; Author; Author; Author

    Titel:
    Development of the Characterization Methods Without Electrothermal Feedback for TES Bolometers for CMB Measurements
    Autor/in / Beteiligte Person: Nishinomiya, Yume ; Kusaka, Akito ; Kiuchi, Kenji ; Terasaki, Tomoki ; Hubmayr, Johannes ; Lee, Adrian ; McCarrick, Heather ; Suzuki, Aritoki ; Westbrook, Benjamin
    Link:
    Zeitschrift: Journal of Low Temperature Physics, Jg. 209 (2022-09-19), S. 1079-1087
    Veröffentlichung: Springer Science and Business Media LLC, 2022
    Medientyp: unknown
    ISSN: 1573-7357 (print) ; 0022-2291 (print)
    DOI: 10.1007/s10909-022-02864-z
    Schlagwort:
    • FOS: Physical sciences
    • General Materials Science
    • Astrophysics - Instrumentation and Methods for Astrophysics
    • Condensed Matter Physics
    • Instrumentation and Methods for Astrophysics (astro-ph.IM)
    • Atomic and Molecular Physics, and Optics
    Sonstiges:
    • Nachgewiesen in: OpenAIRE
    • Rights: OPEN

    Klicken Sie ein Format an und speichern Sie dann die Daten oder geben Sie eine Empfänger-Adresse ein und lassen Sie sich per Email zusenden.

    oder
    oder

    Wählen Sie das für Sie passende Zitationsformat und kopieren Sie es dann in die Zwischenablage, lassen es sich per Mail zusenden oder speichern es als PDF-Datei.

    oder
    oder

    Bitte prüfen Sie, ob die Zitation formal korrekt ist, bevor Sie sie in einer Arbeit verwenden. Benutzen Sie gegebenenfalls den "Exportieren"-Dialog, wenn Sie ein Literaturverwaltungsprogramm verwenden und die Zitat-Angaben selbst formatieren wollen.

    xs 0 - 576
    sm 576 - 768
    md 768 - 992
    lg 992 - 1200
    xl 1200 - 1366
    xxl 1366 -