A nurse‐led implantable loop recorder service is safe and cost effective

Lambiase, Pier D. ; Hunter, Ross J. ; et al.
In: Journal of Cardiovascular Electrophysiology, Jg. 30 (2019-10-09), S. 2900-2906
Online unknown

A nurse‐led implantable loop recorder service is safe and cost effective 

Introduction: Implantable loop recorders (ILR) are predominantly implanted by cardiologists in the catheter laboratory. We developed a nurse‐delivered service for the implantation of LINQ (Medtronic; Minnesota) ILRs in the outpatient setting. This study compared the safety and cost‐effectiveness of the introduction of this nurse‐delivered ILR service with contemporaneous physician‐led procedures. Methods: Consecutive patients undergoing an ILR at our institution between 1st July 2016 and 4th June 2018 were included. Data were prospectively entered into a computerized database, which was retrospectively analyzed. Results: A total of 475 patients underwent ILR implantation, 271 (57%) of these were implanted by physicians in the catheter laboratory and 204 (43%) by nurses in the outpatient setting. Six complications occurred in physician‐implants and two in nurse‐implants (P = .3). Procedural time for physician‐implants (13.4 ± 8.0 minutes) and nurse‐implants (14.2 ± 10.1 minutes) were comparable (P = .98). The procedural cost was estimated as £576.02 for physician‐implants against £279.95 with nurse‐implants, equating to a 57.3% cost reduction. In our center, the total cost of ILR implantation in the catheter laboratory by physicians was £10 513.13 p.a. vs £6661.55 p.a. with a nurse‐delivered model. When overheads for running, cleaning, and maintaining were accounted for, we estimated a saving of £68 685.75 was performed by moving to a nurse‐delivered model for ILR implants. Over 133 catheter laboratory and implanting physician hours were saved and utilized for other more complex procedures. Conclusion: ILR implantation in the outpatient setting by suitably trained nurses is safe and leads to significant financial savings.

Keywords: arrhythmias; implantable loop recorder; syncope

1 INTRODUCTION

An implantable loop recorder (ILR) is a cardiac monitor that is implanted in the subcutaneous tissue of the chest wall. Its ability for long‐term continuous monitoring makes it invaluable in assessing the correlation between symptoms/manifestations such as unexplained syncope and arrhythmias.[[1]]

Recent guidelines[1] have broadened indications for ILR utilization. They have also highlighted their role in the investigation of patients with suspected intermittent arrhythmias, unexplained falls and patients with suspected epilepsy, but ineffective treatment. Furthermore, ILRs are recommended over implantable cardioverter defibrillators (ICD) in patients with inherited cardiomyopathy or channelopathy presenting with recurrent syncope and low risk of sudden cardiac death.[1] Data from the CRYSTAL‐AF[5] study demonstrated the effectiveness of ILRs in diagnosing AF in patients presenting with cryptogenic stroke. The rising number of indications, has therefore been accompanied by an increase in the number of referrals to our center, and, consequently, in the waiting list for this procedure.

Traditionally, ILR implantation has taken place in the catheter laboratory by a cardiologist and is considered a minor surgical procedure. The most recent iteration of these devices are significantly smaller than previous generations, for instance the Reveal LINQ ICM (Medtronic, Minnesota) is 87% smaller than its previous versions, and includes an implanting kit that allows the device to be injected. This new implant technique negates the requirement for a minor surgical procedure,[6] and permits implantation to be undertaken in the outpatient setting.[[7]] We hypothesized that this new implant method would be compatible with a nurse‐delivered implant service in the outpatient setting that would: (a) save physician time, releasing them and the catheter laboratory so other types of more complex procedures could be done, and (b) reduce overall cost.

In the present study, we report the process undertaken to establish this, and evaluate safety and cost‐effectiveness of the service.

2 METHODS

2.1 Patient population

Patients undergoing an ILR implantation during the period 1st July 2016 to 4th June 2018 were included. Data were collected prospectively at the time of implant and during patient follow‐up at 1 month postimplant. Written consent for the procedure was obtained from all patients. This study was prospectively registered and approved by the Clinical Effectiveness Unit at Bart's and the London NHS Trust (ID9883).

Patients were assigned to physician‐led or nurse‐led based on the availability of nurse‐led clinic, and on the type of referral (elective patients were implanted by the nurses, and in‐patients or patients referred directly from A&E were implanted from physicians in the initial months. Later, during the study, spaces were also made available for these patients in the nurse‐led service).

2.2 Nurse‐delivered ILR insertion

The Barts Heart Centre nurse‐delivered ILR implantation service commenced in August 2016. Five nurses undertook a training program to ensure competency in implantation. Procedures were performed in an outpatient clinic room, with a cardiac physiologist assisting and interrogating the device. Briefly, the training scheme was as follows: nurses observed a training video followed by a 10‐minute demonstration on implant method, delivered by company representatives. Implanters then practiced four implants on demonstration kits, and completed a manufacturer's online training module.[9] This was followed by observation of five implants by trained and experienced physician implanters in the catheter laboratory. Five implants were then performed under direct consultant physician supervision. A further five implants were performed with indirect supervision i.e. the supervisor was present in an adjacent clinic room.

Once this was successfully completed, nurses were left with indirect supervision from a consultant cardiologist who will regularly review the progress. The consultant cardiologist was always present in a clinic room in very close proximity, and available to provide help or troubleshooting advice during all ILR implants. Nurses required annual competency testing and updating.

Nurse‐led implants were all done in a minor procedures clinic room (Figure). The implant team consisted of a nurse implanter and a cardiac physiologist. Wound closure was performed using steri‐strips alone. The physiologist role was to provide support during implantation, check ILR function, teach device activator use to the patient and set up CARELINK home monitoring.

jce14206-fig-0001.jpg

2.3 Doctor‐led ILR insertion

ILRs implanted by doctors in the catheter laboratory during the same time period were used as the control group. Wound closure could either be with sutures or steri‐strips alone. Doctor‐led implants, involved the participation of the catheter laboratory team, also including a physiologist, one scrub nurse and one runner nurse to administer sedation if needed.

2.4 Endpoints

The primary endpoint of our study was complication rates. This was defined as the presence of any complication related to the implant, occurring within the first 30 days, and included infection, erosion, migration, and poor R‐wave (<0.2 mV) needing reposition.[10]

Secondary endpoints were: procedure duration, lab/room time utilization, and R‐wave sensing (amplitude). Procedure time was recorded at the time of implant by the physiologist. This was defined from the point of needle‐to‐skin to dressing‐on time. Lab time was also recorded and this was defined as total time spent with patient from the initial entry to leaving the catheter laboratory or outpatient clinic room.

Finally, we performed a cost analysis‐comparison of both approaches taking into account staff needed to be present, lab and material costings, as well as income from additional cases done in the catheter laboratory because of newly created slots resulted from moving ILR cases to nurse‐led clinic.

Cost analysis was done by taking the mean number of ILRs implanted in our centre per annum and the mean lab time used by nurses and doctors. We then calculated total cost by including staff cost and associated cost to determine the overall cost of ILR implantations in the catheter laboratory and outpatient settings. This allowed estimation of total costs, and also cost per procedure. Staff salaries were taken from Royal College of Nursing and NHS Careers.[[11]] We made the assumption that all staff members were full time 45 weeks per year and 30% added on for national insurance/pensions. Associated cost which included cleaning, materials and overheads calculation was based on Kanters et al.[13]

We then compared the two different scenarios and their associated costs: during 1 year all ILR LINQ implants being done by nurses vs all ILR LINQ implants being done by doctors.

For estimating the income from additional cases done in the catheter laboratory because of newly, created slots resulted from moving ILR cases to nurse‐led clinic, we: (a) multiplied the total duration of an ILR case performed in the catheter laboratory by the annual number of ILR implants (ie total time gained per year); (b) Using data for procedural duration from the FIRE & ICE trial[14] we determined how many additional cryoballoon ablation procedures (scenario I) or permanent pacemaker implants (scenario II) could be performed annually; and (c) using Tariff data for pacemaker implants and cryoballoon ablation procedures performed we estimated the additional annual income for both scenarios (additional cryoballoon ablations or pacemaker implants).

2.5 Statistical analysis

Qualitative variables were presented as relative frequencies. Continuous variables were tested for normal distribution using the Kolmogorov–Smirnov test. Normally distributed data were expressed as mean±standard deviation (SD). For categorical variables, the X2 or Fisher exact test, were used to compare the distributions for two or more groups. Non‐paired student T test or Mann‐Whitney U test were used for comparisons of continuous variables. A P < .05 was considered significant. All statistical analyses were performed with SPSS (IBM SPSS Statistics version 22.0).

3 RESULTS

During a 23 months' period (1st July 2016‐4th June 2018) a total of 491 patients underwent an ILR implant. A total of 287 (58%) ILRs were implanted by a doctor in the catheter laboratory and 204 (41%) ILRs were implanted by trained nurses in the outpatient setting.

In 16 of the implants done by doctors, 14 patients had an old ILR explanted and a new one re‐implanted, 1 patient had an ILR implanted following a negative electrophysiology study, 1 patient had an ILR in conjunction with ajmaline test. These cases were excluded from our analysis. Therefore, 475 patients were included in the study with 271 (57%) in the doctors' group and 204 (43%) performed by nurses.

The baseline characteristics of the study population are presented in Table. There was no significant difference in the gender between the study groups However, there were significant differences in the age, pre‐implantation symptoms (syncope, palpitations, others) as well as in the etiology/background of patients (no heart disease, cardiomyopathy, congenital heart disease, others) as shown in Table.

1 Baseline characteristics of patients

Doctor‐led procedureNurse‐led procedureP value
Number of patients271204
Sex (male)121 (45%)91 (45%).742
Age, y50.0 ± 18.554.0 ± 17.5.020
Indication for ILR
Syncope138 (51%)133 (65%).006
Palpitations78 (29%)38 (19%)
Other55 (20%)33 (16%)
Underlying heart condition
None180 (66%)147 (72%).003
CMP25 (9%)27 (13%)
GUCH32 (12%)6 (3%)
Other34 (13%)24 (12%)

1 Note: Values presented as mean ± SD.

  • 2 Abbreviations: CMP, cardiomyopathy (ischemic, hypertrophic/dilated); GUCH, grown up congenital heart disease.
  • 3 a Vertigo/pre‐syncope/dyspnea.
  • 4 b Long QT syndrome/Brugada syndrome/valve disease/postablation/myotonic dystrophy.
3.1 Procedural parameters/features

A significant proportion of patients (12%) who had an ILR implanted by doctors, had intravenous (IV) sedation with Midazolam or Diamorphine compared to none done by nurses (P < .001) (Table).

2 Comparison of procedural characteristics between the study groups

Doctor‐led procedureNurse‐led procedureP value
IV sedation32 (12%)0(0%)<.001
Lab time, min35.5 ± 23.139.0 ± 14.3<.001
Procedure time, min13.4 ± 8.014.2 ± 10.1.986
R‐wave, mV0.50 ± 0.330.45 ± 0.27.567

  • 5 Note: Values presented as mean ± SD.
  • 6 Abbreviations: IV, intravenous; Lab, laboratory; min, minutes.

Moreover, our results showed that nurses used the "operation" room significantly longer than doctors (P < .001). However, the actual procedure of implantation did not differ among the two groups (P = .986) (Table).

Importantly, there was no difference between the two groups with regard to R‐wave values during/after ILR implant for doctors and nurses (0.50 ± 0.33 mV vs 0.45 ± 0.27 mV, P = .567).

3.2 Procedural safety/complications

Overall, a total of eight complications were recorded (1.7%) in the overall cohort. Of those eight cases, six were performed by doctors and two cases were performed by nurses (P = .301) (Table). Infection and erosion were documented only in three patients and one patient, respectively (Table).

3 Complications between the study groups

Doctor‐led procedureNurse‐led procedureP value
Total complications6 (2.2%)2 (1%).301
Infection21
Erosion01
Migration10
Reposition/low R20
Protrusion10

3.3 Current practice and cost analysis of ILR implantation in our centre

Our data demonstrated significant financial savings when ILRs are implanted by nurses in the outpatient setting. Our cost analysis revealed that a single doctor‐led procedure in the catheter laboratory when staff and associated cost are added up, would incur a cost of £576.02 as opposed to £279.95, when done by nurses in the outpatient setting. This would equate to a £296.07 cost difference per procedure

In our centre where we implant 245 ILRs a year, staff cost alone for a doctor‐led procedure would amount to £10 513.13 compared to £6661.55. Adding staff cost to overheads which include running, cleaning and maintenance a doctor‐led catheter laboratory service would cost £130 611.95 vs £61 926.20 if done by nurses in the outpatient setting. This would amount to a grand total of £68 685.75 saved per annum by taking ILR out of the catheter laboratory to an outpatient setting done by nurses (Table).

4 Cost analysis—savings taking ILR out of lab

Catheter laboratory hours saved
In‐LabOut of lab
Number of ILR245245
Time per procedure, min32.8038.50
Time per annum, hours133.93157.21

4 Cost analysis—savings taking ILR out of lab

Staff cost saved
In‐LAB
No. of staffTime per procedure, minSalary per annumSalary per minuteCost per procedureCost per annum
Consultant032.8£84 667£1.02£33.43£ 0
SpR132.8£50 000£0.60£19.74£4836.48
Nurse232.8£27 901£0.34£11.02£2699.90
Physiologist132.8£30 764£0.37£12.15£2976.75
Total: £10 513.13

4 Cost analysis—savings taking ILR out of lab

Out of LAB
No. of staffTime per procedure, minSalary per annumSalary per minuteCost per procedureCost per annum
Consultant038.5£84 667£1.02£39.240
SpR038.5£50 000£0.60£23.170
Nurse138.5£27 901£0.34£12.93£3167.85
Physiologist138.5£30 764£0.37£14.26£3493.70
Total: £6661.55
Total staff cost saved—£3851.58

4 Cost analysis—savings taking ILR out of lab

Associated cost
Based on Kanters et al13
In‐LabOut of Lab
MaterialsEUR 245EUR 245
InstrumentsEUR 100EUR 5
CleaningEUR 57EUR 4
Overhead costEUR 197EUR 30
Total per procedureEUR 599EUR 284
Total in pounds£533.11£252.76
Total associated cost per annum£130 611.95£61 926.20

  • 7 Note: Total associated cost saved—68 685.75.
  • 8 Total cost saved £3851.58 + £68 685.75 = £72 537.33.

Apart from cost savings, our analysis also showed significant savings in regard to lab and implanting physician time. With an average time of 37.3 minutes spent in the catheter laboratory implanting 245 ILRs per year, over 133 catheter laboratory hours can be saved and utilized for other procedures like ablations or pacing.

With an average cryoablation for atrial fibrillation taking up to 120 minutes[14] 133 hours would equate to over 66 additional AF ablations performed. With the tariff for an AF ablation set at £4000,[15] an income of £264 000 could be made before overheads. In the context of pacing, with an average dual chamber pacemaker implantation time of an hour, and a tariff of £2900, an income of £385 700 could be made.[15] This shows how hours saved could potentially generate an additional income stream.

4 DISCUSSION

In the present study, we found that REVEAL LINq implantation by nurses in the outpatient setting is both safe and cost effective.

ILRs are increasingly being used in clinical practice.[16] Their ability for long‐term monitoring allows symptom‐rhythm correlation, making them very attractive for specific indications that were mentioned before.[17] As the indications for ILRs expand, it is likely that their uptake will increase in the future. Therefore, the ability to implant an ILR in a less resource intensive setting by nurses is of importance and has the ability to confer significant long‐term savings. The evolution of ILR LINq implants over the course of this study suggests that the majority, if not all future ILRs, could be successfully implanted in an outpatient setting.

The recent Reveal‐LINq In Office 2 (RIO‐2)[18] randomized control trial compared ILR implants in hospital vs in an office environment. This study demonstrated an excellent safety profile for ILR implant irrespective of the insertion environment/setting.

In our study, we went one step further by introducing a nurse‐led office‐based LINQ‐ILR service. Our results for in‐office implants are comparable with published data, showing similar infection and complication rates.[[18]]

Our data suggest that the nurse‐led approach is a safe alternative to a physician‐led service. This is more likely driven by the ease of implantation of modern day ILRs, where operators can be proficient after a short period of training. The low complication rate among nurses could also be due to the fact that nurses are more likely to adhere to standard operating protocols than doctors. As such, when a nurse becomes more proficient in device implantation, the service can slowly move out of the catheter laboratory completely. This is consistent with the results reported by Kipp et al,[8] where ILR implanted in an ambulatory setting by advanced practise providers had a high success and low complication rate with a single dose of IV antibiotics. Our study differs in that we did not administer IV antibiotics peri‐procedure. In addition, we demonstrated that the outcomes and complications in a nurse‐led service are comparable to a physician‐led service.

Importantly, changing from a catheter laboratory physician‐led service to an office‐based nurse‐led service constitutes cost savings of £296.07 per procedure. In a high‐volume centre like Barts Heart Centre, this resulted in up to £72 000 savings annually. In the current cash‐tight environment existing in the UK health service and worldwide, such savings could enable investment in the development of other areas of service. Further to the cost savings, we demonstrated that an extra 133 catheter laboratory hours could be saved and this could allow more patients to gain access to more EP procedures in a timely manner.

In our study, we found that cases performed by nurses did not require IV sedation. This could be the result of patients feeling more relaxed in the outpatient setting, or because a small minority of patients deemed to be more anxious were likely identified earlier by the referrer and hence done in the catheter laboratory. Having the procedure done in a more office‐like environment with less staff around could also be less intimidating and stressful for the patients.

To the best of our knowledge, our study is the first to assess the above parameters and to suggest a possible introduction of a nurse‐led outpatient ILR implanting service. One of our key strengths is that we are reporting 'real‐world data in a relative large number of patients. All data were collected prospectively and all patients were seen 1 month postimplant, with all complications being logged and recorded.

The main limitation of our study was that this is not a randomized controlled comparison and a retrospective analysis of the two ILR implanting approaches has been performed. However, no differences in complications were observed among the two groups, and it is highly unlikely that minor differences in age, or other non‐accounted characteristics could have a major impact on the observed results. It is also worth noting that this was a prospective non‐randomized study, and there may have been a minimal selection bias resulting in assignment of patients with structurally normal heart mainly to the nurse‐led team. Even though these are less complicated patients, which makes them more suitable to be done in an ambulatory setting, it resulted in nonsignificant difference in procedural time.

5 CONCLUSION

Our data show that ILRs can safely be implanted in the office environment by trained nurses. This operating model resulted in significant cost and time savings and allowed optimal deployment of resources for more complex procedures that would improve service delivery for patients, while creating a stream of additional revenue for the centre.

Footnotes 1 Wei Yao Lim and Nikolaos Papageorgiou contributed equally. 2 Disclosures: RJS has had research agreements and speaker fees from Abbott, Medtronic, Boston Scientific and Biosense Webster. PDL has received educational grants from Medtronic and Boston Scientific. MF has received speaker fees from Biotronik and Medtronic. All other author have no reported relationships relevant to the contents of this paper. REFERENCES Brignole M, Moya A, de Lange FJ, et al. 2018 ESC guidelines for the diagnosis and management of syncope. Eur Heart J. 2018 ; 39 (21): 1883 ‐ 1948. https://doi.org/10.1093/eurheartj/ehy037 Furukawa T, Maggi R, Bertolone C, Fontana D, Brignole M. Additional diagnostic value of very prolonged observation by implantable loop recorder in patients with unexplained syncope. J Cardiovasc Electrophysiol. 2012 ; 23 : 67 ‐ 71. 3 Krahn AD, Klein GJ, Yee R, Hoch JS, Skanes AC. Cost implications of testing strategy in patients with syncope. J Am Coll Cardiol. 2003 ; 42 : 495 ‐ 501. 4 Krahn AD, Klein GJ, Yee R, Takle‐Newhouse T, Norris C. Use of an extended monitoring strategy in patients with problematic syncope. Circulation. 1999 ; 99 : 406 ‐ 410. 5 Sanna T, Diener HC, Passman RS, et al. Cryptogenic stroke and underlying atrial fibrillation. N Engl J Med. 2014 ; 370 (26): 2478 ‐ 2486. https://doi.org/10.1056/NEJMoa1313600 6 REVEAL LINQ Product Specifications—Medtronic. https://www.medtronic.com/content/dam/medtronic‐com/usen/newsroom/media‐resources/media‐kits/reveal‐linq‐insertable‐cardiacmonitor/documents/LINQ%20Fact%20Sheet%20FINAL.pdf 7 Roebuck A, Mercer C, Denman J, et al. Experiences from a non medical, non catheter laboratory implantable loop recorder (ILR) service. Br J Cardiol. 22, 2015 : 36. 8 Kipp R, Young N, Barnett A, et al. Injectable loop recorder implantation in an ambulatory setting by advanced practice providers: analysis of outcomes. Pacing Clin Electrophysiol. 2017 ; 40 (9): 982 ‐ 985. 9 Medtronic. https://www.medtronicacademy.com/learning‐plan/uk‐reveal‐linq‐learning‐plan. Accessed 30 May 2019. Pürerfellner H, Sanders P, Pokushalov E, Di Bacco M, Bergemann T, Dekker LRC. Miniaturized reveal LINQ insertable cardiac monitoring system: first‐in‐human experience. Heart Rhythm. 2015 ; 12 (6): 1113 ‐ 1119. Royal College of Nursing. https://www.rcn.org.uk/employment‐and‐pay/nhs‐conditions‐of‐employment. Accessed 30 May 2019. NHS Health Carers. https://www.healthcareers.nhs.uk. Accessed 30 May 2019. Kanters TA, Wolff C, Boyson D, et al. Cost comparison of two implantable cardiac monitors in two different settings: reveal XT in a catheterization laboratory vs. reveal LINQ in a procedure room. Europace. 2016 ; 18 (6): 919 ‐ 924. https://doi.org/10.1093/europace/euv217 Kuck KH, Brugada J, Fürnkranz A, et al. Cryoballoon or radiofrequency ablation for paroxysmal atrial fibrillation. N Engl J Med. 2016 ; 374 : 2235 ‐ 2245. NHS England and NHS Improvement. 2017/2018 and 2018/2019 National Tariff: currencies and prices. https://improvement.nhs.uk/documents/597/Copy%5fof%5fAnnex%5fA%5f%5fNational%5ftariff%5fworkbook.xlsx. Accessed 30 May 2019. Hindricks G, Camm J, Merkely P, et al. The EHRA white book 2017. Available from: https://www.escardio.org/Sub‐specialty‐communities/European‐Heart‐Rhythm‐Association‐(EHRA)/Research‐andPublications/The‐EHRA‐White‐Books Priori SG, Blomström‐Lundqvist C, Mazzanti A, et al. 2015 ESC Guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: The task force for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death of the European Society of Cardiology (ESC) endorsed by: Association for European Paediatric and Congenital Cardiology (AEPC). Eur Heart J. 2015 ; 36 (36): 2793 ‐ 2867. Rogers JD, Sanders P, Piorkowski C, et al. In‐office insertion of a miniaturized insertable cardiac monitor: results from the reveal LINQ in‐office 2 randomized study. Heart Rhythm. 2017 ; 14 (2): 218 ‐ 224. https://doi.org/10.1016/j.hrtthm Mittal S, Sanders P, Pokushalov E, et al. Safety profile of a miniaturized insertable cardiac monitor: results from two prospective trials. Pacing Clin Electrophysiol. 2015 ; 38 (12): 1464 ‐ 1469. https://doi.org/10.1111/pace.12752

By Wei Yao Lim; Nikolaos Papageorgiou; Shivasankar M. Sukumar; Sophia Alexiou; Neil T. Srinivasan; Christopher Monkhouse; Holly Daw; Helder Caldeira; Helen Harvie; Jincymol Kuriakose; Marco Baca; Syed Y. Ahsan; Anthony W. Chow; Ross J. Hunter; Malcolm Finlay; Pier D. Lambiase; Richard J. Schilling; Mark J. Earley and Rui Providencia

Reported by Author; Author; Author; Author; Author; Author; Author; Author; Author; Author; Author; Author; Author; Author; Author; Author; Author; Author; Author

Titel:
A nurse‐led implantable loop recorder service is safe and cost effective
Autor/in / Beteiligte Person: Lambiase, Pier D. ; Hunter, Ross J. ; Papageorgiou, Nikolaos ; Providência, Rui ; Caldeira, Helder ; Earley, Mark J. ; Ahsan, Syed ; Baca, Marco ; Alexiou, Sophia ; Daw, Holly ; Chow, Anthony ; Monkhouse, Christopher ; Wei Yao Lim ; Shivasankar Murugan Sukumar ; Kuriakose, Jincymol ; Harvie, Helen ; Srinivasan, Neil ; Schilling, Richard J. ; Finlay, Malcolm
Link:
Zeitschrift: Journal of Cardiovascular Electrophysiology, Jg. 30 (2019-10-09), S. 2900-2906
Veröffentlichung: Wiley, 2019
Medientyp: unknown
ISSN: 1540-8167 (print) ; 1045-3873 (print)
DOI: 10.1111/jce.14206
Schlagwort:
  • Adult
  • Male
  • medicine.medical_specialty
  • Databases, Factual
  • Cost-Benefit Analysis
  • Monitoring, Ambulatory
  • 030204 cardiovascular system & hematology
  • Nurse's Role
  • Workflow
  • 03 medical and health sciences
  • Nurse led
  • 0302 clinical medicine
  • Cost Savings
  • Predictive Value of Tests
  • Physiology (medical)
  • Ambulatory Care
  • Implantable loop recorder
  • Outpatient setting
  • medicine
  • Humans
  • 030212 general & internal medicine
  • Physician's Role
  • Aged
  • Retrospective Studies
  • Computerized databases
  • business.industry
  • Financial savings
  • Health Care Costs
  • Middle Aged
  • Catheter
  • Remote Sensing Technology
  • Emergency medicine
  • Female
  • Clinical Competence
  • Cardiology and Cardiovascular Medicine
  • business
Sonstiges:
  • Nachgewiesen in: OpenAIRE
  • Rights: OPEN

Klicken Sie ein Format an und speichern Sie dann die Daten oder geben Sie eine Empfänger-Adresse ein und lassen Sie sich per Email zusenden.

oder
oder

Wählen Sie das für Sie passende Zitationsformat und kopieren Sie es dann in die Zwischenablage, lassen es sich per Mail zusenden oder speichern es als PDF-Datei.

oder
oder

Bitte prüfen Sie, ob die Zitation formal korrekt ist, bevor Sie sie in einer Arbeit verwenden. Benutzen Sie gegebenenfalls den "Exportieren"-Dialog, wenn Sie ein Literaturverwaltungsprogramm verwenden und die Zitat-Angaben selbst formatieren wollen.

xs 0 - 576
sm 576 - 768
md 768 - 992
lg 992 - 1200
xl 1200 - 1366
xxl 1366 -