EMUE-D2-5-TraceabilityFromConformity
In: oai:zenodo.org:4742244; (2021)
unknown
Zugriff:
Measurement traceability is commonly obtained from calibration measurements that provide a result in terms of a single value and its associated uncertainty. However, there are circumstances where instead, the result may consist of a range of possible values. Such circumstances might arise when a result is provided in the form of the output from a conformity decision process, for example as a conformity statement in which a range of acceptable values rather than a specific value is reported. In terms of metrological traceability this style of result provides less information than a specific value, but it may be sufficient to obtain an acceptable target measurement uncertainty for a given application. The standard ISO/IEC 17025 acknowledges this in informative annex A. This example describes how such information might be used to propagate traceability. Files contained in the dataset are: “EMUE 1.2.5.zip”: Zip file containing tex, bib and image files “A125.bib”: Bibliography file “EMUE_A1.2.5.tex”: LaTeX file “EMUE_1_2_5.pdf”: Report - GUM-LPU uncertainty evaluation - importing measurement traceability from a conformity statement “Figure_1_2.m”, “Figure_3.m”: MATLAB scripts used to calculate data for section 3.7
Titel: |
EMUE-D2-5-TraceabilityFromConformity
|
---|---|
Autor/in / Beteiligte Person: | Greenwood, John ; Bošnjaković, Alen ; Karahodžić, Vedran ; Pedone, Paola ; Manta, Fabrizio |
Link: | |
Quelle: | oai:zenodo.org:4742244; (2021) |
Veröffentlichung: | 2021 |
Medientyp: | unknown |
DOI: | 10.5281/zenodo.4742244 |
Schlagwort: |
|
Sonstiges: |
|